MODEL |
7130 |
7132 |
7140 |
7142 |
AC WITHSTAND VOLTAGE |
||||
Output Rating |
5KVAC/20mA |
|||
|
Range |
Resolution |
Accuracy |
|
Output Voltage, KVAC |
0-5.00 |
0.01 |
±(2% of setting + 5V) |
|
Output Frequency |
50Hz/60Hz ±100ppm, User Selection |
|||
Output Waveform |
Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 |
|||
Output Regulation |
± (1% of output + 5V), From no load to full load |
|||
SETTINGS |
||||
Max-Limit AC Current, mA |
0 - 20.00 |
0.01 |
±(2% of setting + 2counts) |
|
Min-Limit AC Current, mA |
0 - 9.999 |
0.001 |
±(2% of setting + 2counts) |
|
Ramp Up Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
|
Ramp Down Time, second |
0 - 999.9 |
0.1 |
|
|
Dwell Time, second |
0, 0.3 - 999.9 (0=continuous) |
0.1 |
|
|
Arc Detection |
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) |
|||
DC WITHSTAND VOLTAGE (7140,7142 ONLY) |
||||
Output Rating |
6KVDC/7500μA |
|||
Output Voltage, KVDC |
0 - 6.00 |
0.01 |
±(2% of setting + 5V) |
|
Output Ripple |
< 5%(6KV/7500μA at Resistive Load) |
|||
SETTINGS |
||||
Max-Limit DC Current, μA |
0 - 7500 |
1 |
±(2% of setting + 2counts) |
|
Min-Limit DC Current, μA |
0 - 999.9 |
0.1 |
±(2% of setting + 2counts) |
|
Ramp Up Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
|
Ramp Down Time, second |
0, 1.0 - 999.9 |
0.1 |
|
|
Dwell Time, second |
0, 0.4 - 999.9(0=continuous) |
0.1 |
|
|
Arc Detection |
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) |
|||
Discharge Time |
≦ 200 msec |
|||
INSULATION RESISTANCE (7132,7142 ONLY) |
||||
Output Rating |
1KVDC/9999MΩ |
|||
Output Voltage, DCV |
30 - 1000 |
10 |
±(2% of setting + 5V) |
|
SETTINGS |
||||
Max-Limit Resistance, MΩ |
0, 1 - 9999 (0=OFF) |
1 |
±(2% of setting+2counts)at 500-1000VDC,1 -1000 MΩ |
|
Min-Limit Resistance, MΩ |
1 - 9999 |
1 |
±(5% of setting+2counts)at 500-1000VDC,1000 -9999 MΩ |
|
|
|
|
±(8% of setting+2counts)at 30-500VDC, 1 -1000 MΩ |
|
Ramp Up Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
|
Ramp Down Time, second |
0, 1.0 - 999.9 |
0.1 |
|
|
Delay Time, second |
0, 0.5 - 999.9(0=continuous) |
0.1 |
|
|
MEASUREMENT |
||||
AC/DC Voltage, KV |
0 - 6.00 |
0.01 |
±(1.5% of reading + 1count) |
|
DC Voltage, V (IR only) |
30 - 1000 |
1 |
±(1.5% of reading + 5V) |
|
AC Current, mA |
0 - 3.500 |
0.001 |
±(2% of reading + 2counts) |
|
|
3.00 - 20.00 |
0.01 |
|
|
DC Current, μA |
0 - 350.0 |
0.1 |
±(2% of reading + 2counts) |
|
DC Current, mA |
0.300 - 3.500 |
0.001 |
|
|
|
3.00 - 7.50 |
0.01 |
|
|
Resistance, , MΩ |
1 - 9999 |
0.001 |
±(2% of setting+2counts)at 500-1000VDC, 1 -1000 MΩ |
|
GENERAL |
||||
Input Voltage AC |
115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A |
|||
PLC Remote Control |
Input : Test , Reset , Memory1.2.3 , Interlock |
|||
|
Output : Pass , Fail, , Processing , Reset-Out |
|||
Interface(Option) |
RS485 Interface |
|||
Memory |
10 memories, 3 steps/memory |
|||
Graphic Display |
128×64 Graphic LCD |
|||
Safety |
Built-in Smart GFI circuit,GFI trip current 450?A max, HV shut down speed: <1mS |
|||
Key Lock |
To prevent unauthorized alteration of the test Parameters |
|||
Calibration |
Build-in software and external calibrated meters |
|||
Verification |
Build-in software driven verification menu to test fault detection circuits |
|||
Alarm Volume Setting |
Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. |
|||
Environment |
0-40℃, 20-80%RH |
|||
Dimension/Net Weight |
215mm(W)×89mm(H)×370mm(D)/ 10Kg |
|||
STANDARD ACCESSORIES |
||||
Power Cord(10A) |
×1 |
|||
Fuses |
×2 (Including a spare contained in the fuse holder) |
|||
Interlock Disable Key(1505) |
×1 |
|||
Hipot Test Lead ,1.5m(1101) |
×1 |
|||
Hipot Return Lead ,1.8m(1102) |
×1 |